Table 2 Accuracy of MgSO4 Delivery

From: Evaluation of a low-cost, low-power syringe pump to deliver magnesium sulfate intravenously to pre-eclamptic women in a Malawian referral hospital

 

Laboratory Performance

Clinical Performance

Mean Error

95% CI

Mean Error

95% CI

Loading Dose

Total Volume Dispensed

−1.16%

[−1.28, −1.04]

−1.48%

[−1.72, −1.25]

Flow Rate

−1.83%

[−2.13, −1.53]

−1.75%

[−2.06, −1.46]

Maintenance Dose

Total Volume Dispensed

−1.43%

[−1.89, −0.97]

−1.13%

[1.30, −0.97]

Flow Rate

−1.67%

[−1.92, −1.42]

−1.11%

[−1.22, −0.10]