Table 5 Agreement between the handheld ultrasound and the reference bigger ultrasound machine for selected Obstetrical ultrasound parameters (N = 460)a

From: Handheld ultrasound to avert maternal and neonatal deaths in 2 regions of the Philippines: an iBuntis® intervention study

Ultrasound parameter In agreement [N, (%)] Not in agreement [N, (%)]
 Fetal presentation 440 (95.7%) 20 (4.3%)b
 Amniotic fluid 458 (99.6%) 2 (0.4%)
 Location 459 (99.8%) 1 (0.2%)
 Fetal number 460 (100.0%) 0 (0.0%)
  1. aHandheld ultrasound: GE VScan and Reference standard ultrasound: GE Logic 5 Premium
  2. bNot in agreement breakdown: Cephalic-Breech (9), Cephalic-Transverse (4), Breech-Transverse (3), Transverse-Breech (2), and Breech-Cephalic (2)